EasyManua.ls Logo

SYSMEX XE-5000 - Page 11

SYSMEX XE-5000
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Document Number: MKT-70-1096 Revision 1 Page 11
SYSMEX XE-5000
TM
9. If the results of the background check is such that the blank value is at
or below the acceptable limit, the READY LED lights and the Main Unit
enters body fluid analysis ready status.
10. Mix the contents of the test tube gently but thoroughly.
Note: Excess mixing can break blood cells in the body fluid,
making it impossible to obtain accurate results.
11. Remove the cap carefully so as not to splatter the sample.
12. Set the test tube into the manual aspiration pipette, and then press the
START SWITCH. Do not remove the test tube while the READY LED is
blinking, sample is being aspirated.
13. After the READY LED turns off (and a short beep sounds two times),
remove the test tube. Once analysis is completed, set Sample to Normal
on the Sample No. setting screen to return to normal mode.
CAUTION: If either the analysis values from body fluid analysis (WBC-BF
and/or RBC-BF) are high, there is the possibility of an influence
on the analysis results for the next sample. If analysis results
for WBC-BF and RBC-BF exceeds specifications, the “Execute
Background Check” Error is displayed, asking the user to per-
form a background check before analyzing the next sample.
14. The results of Body Fluid Mode are displayed on the menu screen of
the Main Unit LCD screen. The results are also displayed on the sample
explorer and the data browser screen Body Fluid Tab on the information
processing unit (IPU)
Analysis samples analyzed in body fluid analysis mode are marked with
an “F” mark to the left of the sample No. on the Main Unit LCD screen.
“F” is displayed in the analysis mode display column of the sample
explorer and data browser screens.
Body Fluid Mode
Explorer Screen “F” and Red “F”
Body Fluid Tab on Browser

Other manuals for SYSMEX XE-5000