Document Number: MKT-70-1096 Revision 1 Page 6
SYSMEX XE-5000
TM
SAMPLER (AUTO) MODE with BARCODES
1. Verify XE Main Unit ‘Ready’ LED is on and ‘Manual’ is displayed on
Main Unit.
2. Place barcoded sample(s) in sample rack(s) with barcodes facing front of
rack.
3. Place rack on right side of Sampler Unit.
4. Press [SAMPLER] on the Main Unit keypad.
5. Select test to be performed for instruments that are not interfaced.
6. Press [Start] on the Main Unit menu or press [SAMPLER] again.
Sample Number and Discrete test will be transmitted from LIS.
Note: If XE is in sleep status, XE will automatically return to Ready when the
Sampler Mode is started. Wait for Ready to appear on the screen.
Sampler Screen
SAMPLER (AUTO) MODE without BARCODES
(Sequential numbering)
1. Set the ID Barcode reader to off on the IPU.
• From the IPU Menu Window area, select ‘Controller’, ‘Settings’ and
‘ID Reader’. Deselect the ID Reader to turn it off. Click [APPLY].
• From the Menu bar, click [Settings] and [Analysis Ordering].
Select ‘Rack No./Tube Pos.’
2. Press [SAMPLER] on Main Unit and sampler setting screen appears.
3. Enter the first sample number (up to 15 alphanumeric characters) in the
rack and the tube position. (ex. 101) Numbers are assigned sequentially
thereafter.
4. Press [↓] to select Discrete test (1 to 7).
5. Place rack on Right Rack Pool and press [START] on the Function Menu,
or press [SAMPLER] again.
INTERRUPT AUTO MODE FOR STAT
1. Press [SAMPLER] on Main Unit LED to interrupt Sampler Operation.
2. ‘S-READY’ appears at the upper left of Main Unit. Ready LED is on.
3. Analyze STAT using the Manual or Capillary Mode.
4. After STAT is completed, ‘S-READY’ appears at the upper left of the Main
Unit screen. Press [SAMPLER] and XE resumes Sampler Mode Analysis.
Sample Processing
XE-5000 MANUAL OPEN MODE
1. Verify XE green ‘Ready’ LED is on.
2. Press [MANUAL] on XE Main Unit LCD panel.
• Enter Sample No. or Name. (13 characters). For
Name, press ‘Num’ key 1, 2 or 3 times to display
desired character (NUM, ALP, alp).
• Press [↓] and [1] to select ‘Manual Mode’.
• Press [↓] to select Discrete test. Press number
(1 to 7) to select desired test.
• Press [ENTER].
3. Mix sample prior to analysis. Uncap tube.
4. Hold uncapped tube under aspiration pipette. Press
the START SWITCH.
5. Remove tube when ‘Ready’ LED stops blinking and
‘beep-beep’ sound is heard. No need to wipe pipette.
6. Results display after completion of analysis.
Manual Aspiration/START SWITCH
Manual Screen