8. CALIBRATION
Turn on the scale.
Press key during self checking, display will be show f 1 unt
Press key until display will be show tech
Note: Before enter the tech menu, press calibration switch, which is
locating below the scale
Press and keys, display will be show p 1 lin
8.1 Linear Calibration li near
Press key to enter calibration, display will be shown pi n
Press and keys, display will be show kg or lb, press key
to select the calibration unit kilograms or pounds, press key ,display will be
show Load0.
*Note: Ensure the pan is empty
After stable and zero indicator on, display will be show Load1
Place the first calibration mass weight on the pan.
After stable and zero indicator on, display will be show Load 2
Add again the second calibration mass weight on the pan.
After stable and zero indicator on, display will be show Load 3
Add again the third calibration mass weight on the pan.
After stable and zero indicator on, display will be show Load 4
Add again the fourth calibration mass weight on the pan, ( *Note: Calibration
mass weight should be full capacity of the scale)
After stable and zero indicator on, display will be show load 0 Remove the
calibration mass weight from the pan
*Note: Ensure the pan is empty
After stable and zero indicator on, display will be show Load 4
Place the calibration mass weight on the pan, ( *Note: Calibration mass
weight should be full capacity of the scale)
After stable and zero indicator on, display will be show Load 3
Remove the fourth calibration mass weight from the pan