Questions and Answers

  • J
    jeanne80Aug 18, 2025
    What to do if my T Scale Scales weight value is incorrect?
    • T
      Tara BrewerAug 18, 2025
      If your T Scale Scales show an incorrect weight value, it could be due to a calibration error, linearity error, using inaccurate weights during linearity calibration, the balance not being leveled, materials between the sample and cover, or the wrong weighing unit being displayed. Try calibrating again, carefully checking the test weight, scale stability, and the weighing unit. If needed, perform a linear calibration. Also, make sure the pan is correctly installed and the scale is in an appropriate location.
  • P
    pwaltersAug 20, 2025
    How to fix T Scale IP30 when the system hangs?
    • C
      Corey ParkerAug 20, 2025
      If your T Scale Scales system hangs, it's likely due to a software error. To resolve this, hold the ON/OFF key for 6 seconds to restart the scale.
  • W
    William FlemingAug 23, 2025
    What to do if my T Scale Scales cannot turn on?
    • E
      Eric ColonAug 23, 2025
      If your T Scale Scales won't turn on, the issue might be that the main power is off, the power supply isn't plugged in, the power supply is faulty, or the display is off. Check that the power switch is turned on and the power LED is working. Ensure the voltage going to the scale matches the power supply labels on the power module or scale.
  • H
    Hannah BlackAug 26, 2025
    Why is my T Scale Scales display unstable?
    • K
      Kevin FigueroaAug 26, 2025
      An unstable display on your T Scale Scales can be caused by air variation, the pan contacting foreign materials or being incorrectly installed, the sample moving, table vibrations, dramatic temperature changes, or a faulty power supply/battery. Ensure the scale is on a stable surface in an appropriate location and that the power supply is correct for the scale.
  • T
    Theresa AlvarezAug 28, 2025
    What to do if my T Scale IP30 Scales LCD is off but the power LED is on?
    • N
      Nicole WalterAug 28, 2025
      If the LCD on your T Scale Scales is off but the power LED is on, the scale may be in sleep mode or the system may have hung. Try pressing the ON/OFF key to return to working mode, or hold the ON/OFF key for 6 seconds to restart the scale. Alternatively, disconnect the battery and charger, then reconnect them and turn on the power.
  • A
    andersonmelanieAug 30, 2025
    What to do if my T Scale IP30 Scales touch screen is incorrect?
    • R
      Richard HardinAug 31, 2025
      If the touch screen on your T Scale Scales is not working correctly, it may need correction, or the housing might be over-tightened. Try connecting a mouse and using it to enter the touch screen correction to calibrate the touch screen. If the housing is over-tightened, loosen the housing screws.
  • U
    uvanceSep 2, 2025
    How to fix T Scale IP30 not linear?
    • M
      Mrs. Kayla KellySep 2, 2025
      If your T Scale Scales are not linear, it could be due to the overload stopper touching something, damaged load cell or mechanics, or a damaged A/D. Verify there are no objects under the pan or load cell, transporting lock is removed and pan installation is correct. Re-calibrate and perform a linear calibration using ideal standard weights.
  • K
    kaylamaldonadoSep 4, 2025
    What to do if my T Scale Scales cannot use full capacity?
    • J
      Juan WarrenSep 4, 2025
      If your T Scale Scales cannot use its full capacity, the overload stopper might be touching the pan support or the bottom of the load cell, the transportation lock might not be removed, there could be an A/D circuit problem, parameters might be set incorrectly, or the load cell/mechanics could be damaged. Check for any objects under the pan or load cell, verify the transporting lock is removed, and ensure the pan is correctly installed. Also, check the weighing units being used.
  • K
    khardingSep 7, 2025
    What causes off center loading error in T Scale IP30?
    • D
      David BentonSep 7, 2025
      Off-center loading errors in T Scale Scales can be caused by adjustment mechanism issues, incorrect overload stops, or a damaged load cell. Verify any objects under the pan or load cell, transporting lock, and verify pan installation. Ensure load cell fixed correctly.
  • E
    Erin MosesSep 9, 2025
    What to do if my T Scale IP30 battery is low?
    • M
      maryobrienSep 9, 2025
      If your T Scale Scales battery is low, verify that the batteries are rechargeable types and that the charging voltage is correct. Plug in the AC adapter to charge the battery. If necessary, replace the battery.