M503 Series Wheelchair Scales Technical Manual
- 14 -
8. CALIBRATION
Turn on the scale.
Press key during normal weighing, display will show f1 off
Press key until display will show tech
Press key to confirm display will show PIN
Press and keys, display will show p1 spd
Press key display show p2 cal
press key, Display will show desc, press key several times
until display show cal, press key into calibration, display will show
UNLOAD,
When scale get stable, Press key, display will show the calibration
value, you can change the value by pressing and key, then
press
key to comfirm.
display will show load
Place the calibration weight on the platform, press key after stable,
display will show pass.
Calibration will be finished, scale will start self test.
After three times, scale will be calibrate automatically and will start self test.
Note: Incase display will show any error message or incorrect measurement,
repeat the calibration again.