M503 Series Wheelchair Scales Technical Manual
8. CALIBRATION
Turn on the scale.
Press key during normal weighing, display will show “f1 off”
Press key until display will show “tech”
Press key to confirm display will show “PIN ”
Press and keys, display will show “p1 spd”
Press key , display show ”p2 cal”
press key, Display will show “desc”, press key several times
until display show ”cal”, press key into calibration, display will show
“UNLOAD”,
When scale get stable, Press key, display will show the calibration
value, you can change the value by pressing and key, then
press
key to comfirm.
display will show “load”
Place the calibration weight on the platform, press key after stable,
display will show “pass”.
Calibration will be finished, scale will start self test.
After three times, scale will be calibrate automatically and will start self test.
Note: Incase display will show any error message or incorrect measurement,
repeat the calibration again.