Discovery TGA™ Getting Started Guide Page 30
3 Using tweezers, compress the foam and rotate it 90 degrees to eliminate contact with the beam. Gently
remove the foam inserts from the sample and tare sides, being careful not to touch the balance. Refer to
the figure below.
Figure 12 Removing the foam inserts.
4 Put the balance cover back into place, tightening the screws to secure it. See Figure 11
above.
Carefully remove this foam