EasyManua.ls Logo

TCi MFC A - PCB Calibration Process

Default Icon
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MFC Kit Instructions 5.0 PQconnect Option
33
PCB Calibration Process
If the user needs to calibrate the PCB than they will need to follow the steps below. Note: an RS485
to USB converter, a windows computer, and PQvision will be needed for the calibration. Before
energizing the filter, the user should connect the RS485 to USB converter to the PCB.
Power will need to be energized to the PCB for the calibration process.
Step 1. Energize unit
Step 2: Open PQvision desktop application
Step 3: Connect to the device by selecting the appropriate “COM port”
Step 4. Go to “Menu” and select Tech access
Figure 17: PQvision Main Menu
Step 5: Go to “Menu” and select “Calibrate
Figure 18: Calibration Screen
Step 6: Select “Default Non-Cal Parameters
Step 7: Select “Start Calibration
o Note: Calibration will take about 1-2 minutes
Step 8: After the calibration is complete, select Save Parameters”