Programmer`s manual
EZE30
www.tecsis.de BE 948 a Annex A Page 21
DX Half or full duplex this command select half or full duplex communication.
Parameter = 0 select half duplex communication and parameter = 1 select full duplex
Communication - use half duplex setting when using two wire RS485.
3.11 ANALOGUE OUTPUT SET-UP COMMANDS AL, AH, AA
AL Set the analogue low level. This command defines the weight value for 4mA analogue output. Issuing the AL
command without any parameters will return the current low level value in the form L+00000. Factory default setting: 0.
AH Set the analogue high level. This command defines the weight value for 20mA analogue output. Issuing the AH
command without any parameters will return the current high level value in the form H+10000. Factory default setting:
10000.
AA Sets the source which will be used by the analogue output. Issuing the AA command without any parameters returns
the current value in the form A+00001.
Factory default setting: 1.
NOTE: Remember to store the settings using the AS command before turning off the
power.
An = 0
Gross weight as analogue output source
An = 1
Net weight as analogue output source