CAUTION
č
CAUTION
č
To
prevent
possible
probe
damage,
be
sure
only
the
sealed
portion
of
the
probe
is
immersed
(see
Fig.
10).
Put
the
probe
tip
into
the
water,
avoiding
the
side
or
bottom
of
the
container.
Wait
for
the
readout
to
stabilize,
indicating
the
probe
has
reached
the
water
temperature.
The
readout
should
be
—2°C
to
2°C.
There
should
be
ice
remaining
after
the
test
to
verify
that
inserting
the
probe
did
not
raise
the
water
temperature.
@
475A/DM44
Operators
High
Temperature
Bring
water
to
a
slow
boil
(to
prevent
splattering).
Put
the
probe
tip
into
the
water,
avoiding
the
side
or
bottom
of
the
container.
Wait
for
the
readout
to
stabilize,
indicating
the
probe
has
reached
the
water
temperature.
The
readout
should
be
between
98°C
and
102°C
far
clean
water
at
sea
level,
TIME
AND
1/TIME
See
DM44
DELAYED
OR
MIXED
SWEEP
TIME
MEASUREMENTS
in
this
manual.
27