EasyManua.ls Logo

Tektronix 475A - Frequencv Measurements; Time Duration

Tektronix 475A
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ju
a
ba
kr
ka
s
SE
a
HORIZONTAL
DISTANCE
Fig,
18.
Time
duration.
Measure
the
horizontal
distance
between
the
time
measurement
points.
Multiply
the
distance
measured
by
the
setting
of
the
TIME/DIV
switch.
If
sweep
magnification
is
used,
divide
this
answer
by
10.
EXAMPLE:
The
distance
between
the
time
measure-
ment
points
is
8.3
divisions
(see
Fig.
18)
and
the
TIME/DIV
switch
is
set
to
2
ms
with
the
magnifier
off.
Using
the
formula:
horizontal
i
Time
|
Å
Time/Div
A
==
distance
X
i
Duration
Nan
setting
(divisions)
magnification
Substitute
the
given
values:
Time
Duration
FREQUENCY
MEASUREMENTS
Time
Duration
measurements
can
be
used
to
determine
the
frequency
of
a
recurrent
signal
as
follows:
1.
Measure
one
cycle
of
the
waveform
as
described
in
Time-Duration
Measurements.
2.
Take
the
reciprocal
of
the
time
duration
to
deter-
mine
the
frequency.
EXAMPLE:
The
frequency
of
the
signal
shown
in
Fig.
18,
which
has
a
time
duration
of
16.6
milliseconds
is:
1
1
nene
time
duration
u
16.6
ms
=
60
hertz
38