RISETIME
(DM44)
Set
controls
as
follows:
FUNCTION
TIME
HORIZ
DISPLAY
A
INTEN
B
SOURCE
STARTS
AFTER
DELAV
B
TIME/DIV
3
or
4
positions
more
cw
than
A
TIME/DIV
A
TIME
To
move
the
time-
measurement
point
to
the
right
of
the
reference
point
Set
the
A
TIME/DIV
switch
to
a
setting
that
displays
several
events.
Set
the
VOLTS/DIV
and
VAR
VOLTS/DIV
so
the
amplitude
of
the
displav
is
exactly
5
divisions.
Vertically
position
the
display
so
the
bottom
touches
the
0%
graticule
line
and
the
top
touches
the
100%
graticule
line.
60
475A/DM44
Operators
Adjust
the
DELAY
TIME
POSITION
control
to
move
the
reference
point
to
the
10%
graticule
line
(see
Fig.
32
Point
A).
Adjust
the
A
TIME
control
to
move
the
time
measure-
ment
point
to
the
90%
graticule
line
(see
Fig.
32,
Point
B).
The
Readout
and
Scale
Factor
Lamps
indicate
the
risetime.
INTENSIFIED
ZONES
i
i
AMPLITUDE
TT
|
HORIZONTAL
|
>
DISTANCE
(465/DM-0-11)2039-29
Fig.
32.
Risetime.