OPERATING VOLTAGE
~
This
instrument
may
be
damaged
if
operated
with
the line voltage
selector
switch
or the
regulating
range
selector
set
for
the
wrong
applied
line voltage
or
if
the
wrong
line fuse is
used
.
Line Voltage Selection
This
instrument
operates from
either
a 115-volt
or
a
230-
volt
nominal
line
voltage
source, 48 to 440 hertz.
To
convert
from
one
nominal
line
voltage
range to the other,
move the Line
Voltage
Selector
switch
(located
on
side
panel) to
indicate
the
correct
nominal
voltage. A 115-to-
230
volt
adapter
may be required
for
the
line-cord
plug
. Be
sure to use the
correct
line fuse for the line
voltage
selected (see
Table
1)
.
Regulating Range Selection
The
Regulating Range
Selector
assembly (located on
the rear
panel) is set
for
one
of
the line voltage ranges
shown
in Table 1. It also
contains
the
line fuse
for
overload
protection.
To
change
the
regulating
range:
1.
Disconnect
the
instrument
from the
power
source
.
2. Loosen the
two
captive screws
that
hold
the cover
on the
selector assembly; then
pull
to
remove the
cover
.
3.
Pullout
the range
selector
switch
bar (see Fig. 1).
Select a range from Table 1
which
is
centered
about
the
average
line voltage. Slide the bar to the desired
position
and
plug
it in. Push the
cover
on and
tighten
the
screws
.
TABLE
1
Regulating Ranges
Regulating Range
Range
Selector
115-Volts. 230-Volts
Switch Position
Nominal Nominal
LO
(Switch bar
in
lower
holes)
99 to
121
volts
198
to
242 volts
M (Switch bar
in
middle
holes) 104
to
126
volts
207
to
253
volts
HI (Switch
bar
in
upper
holes)
108
to
132 volts 216
to
264 volts
Fuse Size
1.5 A
3AG
0.75 A
3AG
Fast-blow Fast-blow
2
475A/DM44 Operators
@
I
I
I
I
I
I
I
I
I