EasyManua.ls Logo

Tektronix 503

Tektronix 503
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Check Type 503
i
- J
t
Trace
seperation
easurement
, , . m
Fig. 5-1. Typical display, am plifier balance check (DC common
mode rejection ratio) vertical amplifier; 1 mV/CM sensitivity.
f. Change the Standard Amplitude Calibrator output to
0.5 volts, and the HORIZONTAL SENSITIVITY switch to 1
mV/CM, and repeat step (d).
g. CheckThe horizontal amplitude of the display should
not exceed 5 cm anywhere in the graticule area.
h. Change the HORIZONTAL SENSITIVITY switch to .5
VOLTS/CM, and the Standard Amplitude Calibrator to 50
volts, and repeat step (d).
i. CheckThe horizontal amplitude of the display should
not exceed 2 cm anywhere in the graticule are, (50:1 CMRR).
j. Change the HORIZONTAL SENSITIVITY switch to 5
VOLTS/CM and the Calibrator output to 100 volts; repeat
step (d).
k. CheckThe horizontal amplitude of the display should
not exceed 4 mm anywhere in the graticule area.
13. Check Compression or Expansion Vertical
a. Requirement— Display amplitude change not more than
±1.5 mm with a 2-cm display.
b. Change the Standard Amplitude Calibrator output to
.5 volts.
c. Set the VERTICAL + INPUT switch to DC, and the
HORIZONTAL + and -INPU T to GND.
d. Move the Calibrator signal from the HORIZONTAL
+ INPUT connector to the VERTICAL +INPUT connector,
and set the VERTICAL SENSITIVITY switch to .2 VOLTS/CM.
e. The display should consist of two dots spaced verti
cally; center the display vertically and horizontally.
f. With the VERTICAL VARIABLE control, adjust the dot
spacing for 2 cm exactly, and move the display to the top
2 cm of the graticule.
g. Check Dot spacing 2 cm ±1 .5 mm.
h. Move the display to the bottom 2 cm of the graticule.
i. Check Dot spacing 2 cm ± 1.5 mm.
j. Return the VARIABLE control to CALIBRATED.
14. Check A m plifier Balance Vertical (DC
Common M ode Rejection)
a. Requirement— Rejection ratios and deflection toler
ances as listed in Table 5-1, with specified input signals and
deflection sensitivities.
b. Set the VERTICAL INPUT switch to DC, and move
the Calibrator signal from the HORIZONTAL INPUT con
nector to the VERTICAL INPUT connector.
c. Change the Standard Amplitude Calibrator to 5 volts.
d. Change the HORIZONTAL DISPLAY switch to SWEEP
NORMAL (X I). The INTENSITY may have to be increased
to view the display.
e. With the VERTICAL POSITION control, move the dis
play to the top line of the graticule, and then to the bottom
line of the graticule.
f. Check— Deflections as indicated in Table 5-1. See
Fig. 5-1 for a typical display.
TABLE
5-1
Input Deflection
Rejection
'Maximum
Deflection
(trace
Signal
Sensitivity Ratio separation)
5 volts .2 VOLTS/CM 100:1
2.5 mm
0.5 volts
1 mV/CM
100:1
5 cm
50 volts .5 VOLTS/CM 1
50:1 2 cm
100 volts 5 VOLTS/CM |
50:1
4 mm
'The degree of
trace separation should be checked by positioning
the display over the entire graticule area.
15. Check Attenuator Accuracy Vertical
+ Input and Input
a. Requirement— Vertical deflection accuracy within ±3 % .
b. Set the VERTICAL -INP UT switch to GND.
c. Change the Standard Amplitude Calibrator output to
5 mV, and change the VERTICAL SENSITIVITY switch to 1
mV/CM.
d. Check Vertical deflection of 5 cm ±1 .5 mm.
e. Set the Standard Amplitude Calibrator output and
VERTICAL SENSITIVITY switch as indicated in Table 5-2,
checking for the deflections listed.
f. Change the VERTICAL +INPUT switch to GND, and
change the VERTICAL INPUT switch to DC.
g. Repeat step (e).
16. Check Attenuator Accuracy Horizontal
+ Input and Input
a. Requirement— Horizontal deflection accuracy within
±3% .
b. Move the two 50-ohm cables from the VERTICAL + IN
PUT and -INPUT connectors to the HORIZONTAL -f INPUT
and -INPU T connectors.
5-4
©

Table of Contents

Related product manuals