EasyManuals Logo
Home>Tektronix>Test Equipment>503

Tektronix 503 User Manual

Tektronix 503
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #70 background imageLoading...
Page #70 background image
Calibration Type 503
q. Check— Trace separation (measured on the graticule
center vertical line) at the widest part of the ellipse not more
than 1 mm (1 °).
r. Adjust C368 for minimum trace separation; adjust as
necessary for minimum trace separation at 350 kHz and
450 kHz.
NOTE
A compromise adjustment of C368 may be neces
sary to achieve minimum trace separation at both
350 kHz and 450 kHz.
s. Move both signal cables to the HORIZONTAL and
VERTICAL + INPUT connectors, and change both + INPUT
switches to DC, and both INPUT switches to GND.
t. Check—Trace separation not more than 1 mm.
u. Adjust C468 for minimum trace separation; adjust as
necessary for minimum trace separation at 350 kHz and
450 kHz.
NOTE
Slight re-adjustment of C419 may be required to
obtain minimum trace separation.
v. Change the VERTICAL and HORIZONTAL SENSITIVITY
switches to 1 mV/CM.
w. Adjust the generator output for an X and Y deflection
of 6 cm at 450 kHz, and center the trace.
x. Check— Trace separation not more than 1 mm.
y. Adjust C356 for minimum trace separation; adjust as
necessary for minimum trace separation at 350 kHz and
450 kHz.
z. Change the signal cables to the INPUT connectors,
set the INPUT switches to DC, and the + INPUT switches
to GND.
aa. Check—Trace separation not more than 1 mm.
ab. Adjust C456 for minimum trace separation; adjust
as necessary for minimum trace separation at 350 kHz and
450 kHz.
ac. Check the remaining attenuator positions for both
the HORIZONTAL and VERTICAL Amplifiers, referring to
Table 6-5 for attenuator settings, generator frequencies,
and trace separation tolerances.
NOTE
In all cases, adjust the display amplitude to 6 cm
for the checks in Table 6-5.
ad. Interaction—The adjustments to C356, C456, and C368,
interact. Changes to the adjustments of one of the
capacitors m ay require adjustment of the others.
ae. Re-check the amplifier bandwidth, using steps
31
(a)
through (o) as a guide. This completes the phasing recali
bration and check.
TABLE 6-5
SENSITIVITY
HORIZONTAL
and
VERTICAL
Generator
Frequency
Maximum
Trace
Separation
1 mV/CM
50 kHz
1 mm
350 kHz
450 kHz
2 mV/CM 50 kHz
1 mm
350 kHz
450 kHz
5 mV/CM 50 kHz 1 mm
350 kHz
450 kHz
10 mV/CM
50 kHz
1 mm
350 kHz
450 kHz
20 mV/CM 350 kHz
1 mm
50 kHz
450 kHz
.1 VOLTS/CM
50 kHz 1 mm
350 kHz
450 kHz
.2 VOLTS/CM
50 kHz
1 mm
______
/
350 kHz
450 kHz
32. Check Common Mode Rejection Ratio
a. Set the HORIZONTAL + and INPUT switches to
GND.
b. Set the VERTICAL + INPUT to DC, and the -IN PU T
to GND.
c. Set the constant-amplitude signal generator frequency
to 50 kHz, and connect the 50-ohm cables to the VERTI
CAL + and INPUT connectors.
d. Set the VERTICAL SENSITIVITY to 1 VOLTS/CM, and
adjust the generator output for a trace 4 cm in amplitude.
e. Change the VERTICAL INPUT switch to DC, and
change the VERTICAL SENSITIVITY switch to .2 VOLTS/
CM.
f. W illi Hie VERTICAL POSITION control, move fTie dis
play to the top, and then to the bottom of the graticule.
g. CheckVertical deflection (trace separation) not more
than 2 mm anywhere in the graticule area.
©
6-23

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 503 and is the answer not in the manual?

Tektronix 503 Specifications

General IconGeneral
BrandTektronix
Model503
CategoryTest Equipment
LanguageEnglish

Related product manuals