7.
Adjust
Up
Peak
(
Loop
1
,
A60
)
a
.
Press
the
following
buttons
in
the
order
listed
:
(
SETUPS
)
RECALL
1
(
PARAMETER
)
FREQ
2
1
ENTER
b
.
Connect
a
1X
probe
from
the
differential
comparator
oscilloscope
to
the
center
conductor
of
P1550
.
The
center
conductor
can
be
accessed
on
the
front
of
the
Interconnect
board
.
Connect
the
probe
ground
lead
to
analog
ground
on
P1550
.
Note
the
amplitude
of
the
negative
peak
.
c
.
ADJUST
—
R1645
,
Up
Peak
,
for
a
positive
peak
am
plitude
equal
to
within
1
mV
of
the
value
noted
.
Set
Power
off
and
install
shields
.
Set
Power
on
.
Restores
frequency
and
symmetry
DACs
to
de
fault
or
previous
settings
.
d
.
Leave
these
connections
for
the
next
step
.
8.
Adjust
Clamp
Offset
and
Clamp
Gain
(
Loop
1
,
A60
)
a
.
Make
certain
the
oscilloscope
is
connected
to
P1550
.
b
.
Press
the
following
buttons
in
the
order
listed
:
(
MODE
)
GATE
(
PARAMETER
)
FREQ
1
EEX
3
ENTER
(
PARAMETER
)
PHASE
9
ENTER
c
.
Connect
a
+1
V
to
-1
V
1
kHz
squarewave
to
the
TRIG
/
GATE
IN
connector
.
REV
JUL
1982
d
.
ADJUST
-
R1151
,
Clamp
Offset
,
until
the
baseline
level
is
within
1
mV
of
the
negative
triangle
peaks
.
Internal
Adjustment
Procedure
-
FG
5010
e
.
Press
the
following
buttons
in
the
order
listed
:
(
PARAMETER
)
PHASE
9
ENTER
f
.
ADJUST
-
R1152
,
Clamp
Gain
,
until
the
baseline
level
is
within
1
mV
of
the
positive
triangle
peaks
.
g
.
Remove
the
signal
from
the
TRIG
/
GATE
IN
and
the
oscilloscope
probe
for
the
next
step
.
9.
Adjust
Phase
Gain
and
Phase
Offset
(
Preliminary
Setting
)(
Loop
1
,
A60
)
a
.
Press
the
following
buttons
in
the
order
listed
:
(
PARAMETER
)
PHASE
9
ENTER
b
.
Connect
the
dvm
+
lead
to
U1042
pin
8
,
-lead
to
analog
ground
.
c
.
CHECK
-
for
a
dvm
reading
from
-1.965
V
to
-1.955
V.
d
.
ADJUST
R1052
,
Phase
Offset
,
for
a
dvm
reading
from
1.965
V
to
1.955
V.
www
e
.
Press
the
following
buttons
in
the
order
listed
:
(
PARAMETER
)
PHASE
9
ENTER
f
.
CHECK
-
for
a
dvm
reading
from
-7.42
V
to
-
7.40
V.
g
.
ADJUST
-
R1051
,
Phase
Gain
,
for
a
dvm
reading
from
7.42
V
to
-7.40
V.
h
.
Remove
these
connections
for
the
next
step
.
6-3