Model 2450
SourceMeter® Instrument User's Manual Section 7: Measuring I-
V characteristics of FETs
2450-900-01 Rev. E / August 2019 7-9
Set up the application using SCPI commands in a linear sweep
In this example, the gate voltage steps from 2 V to 6 V in 1 V steps, the drain voltage sweeps from
0 V to 5 V in 51 steps, and the drain current is measured. The application retrieves each of 51
readings and source values from defbuffer1 and saves them in the arrays vds and ids,
respectively.
You send commands to either the sweeper (SMU 2) or the stepper (SMU 1). In the table, the
commands for the sweeper are shown with a light gray background. The commands for the stepper
are shown with a darker gray background. The light-brown shaded code represents pseudocode that
will vary depending on the programming environment you use. Each bulleted item in the Description
column describes a single line of code in the Commands column.
Send the following commands for this application:
SMU 1,
SMU 2, or
pseudocode
Commands Description
SMU 1
*RST
:SENS:FUNC "CURR"
:SENS:CURR:RANG:AUTO ON
:ROUT:TERM REAR
:SOUR:FUNC VOLT
:SOUR:VOLT:RANG 20
Reset the instrument.
Set to measure current.
Set to measure with autorange
enabled.
Set to use rear terminals.
Set to source voltage.
Set to 20 V source range.
SMU 2
*RST
:SENS:FUNC "CURR"
:SENS:CURR:RANG:AUTO ON
:ROUT:TERM REAR
:SOUR:FUNC VOLT
:SOUR:VOLT:RANG 20
:SOUR:VOLT:ILIM 1
:SOUR:SWE:VOLT:LIN 0, 5, 51, 0.01
Reset the instrument.
Set to measure current.
Set to measure with autorange
enabled.
Set to use the rear terminals.
Set to source voltage.
Set to the 20 V source range.
Set the source limit to 1 A.
Set up a linear sweep from 0 V to 5 V
in 51 steps, with a delay of 10 ms.
SMU 1
:OUTP ON
Turn on the output.
Pseudocode
vds = []
ids = []
for i = 2, 5 do:
Create an empty array to hold
measured voltage values.
Create an empty array to hold
measured current values.
Set up a for loop from 2 to 5
(inclusive).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2450 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2450 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2450
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals