Installation
P6316 Digital Probe
Threshold Accuracy: ±(100 mV + 3% of threshold)
Maximum signal swing: 20 V
peak-to-peak
centered around the threshold voltage
Minimum signal swing: 500 mV
peak-to-peak
Maximum nondestructive input signal to probe: 40 V
peak-to-peak
Input resistance: 101 k
Input capacitance: 8.0 pF
Temperature:
Operating: 0 °C to +50 °C (+32 °F to +122 °F)
Nonoperating: -40 °C to +71 °C (-40 °F to +160 °F)
Altitude:
Operating: 3,0 00 m (9,843 ft) maximum
Nonoperating: 12,00 0 m (39,37 0 ft) maximum
Pollution Deg ree: 2, Indoor use only