Specifications
Table 4: Trigger specifications (cont.)
For signals having rise and fall times 10 ns.
Source
Range
Any input channel ±0.20 div
Aux In ( External) ±(10% of setting + 25 mV)
Trigger level accuracy, DC coupled,
typical
Line Not applicable
Trigger holdoff range
20 ns minimum to 8 s m aximum
Maximum serial trigger bits 128 bits
Standard serial bus interface triggering
I
2
C Address Triggering: 7 and 10 bit user specified address, as well as General Call, START
byte, HS-mode, EEPROM, and CBUS
Data Trigger: 1 to 5 bytes of user specified data
Trigger On: S tart, Repeated Start, Stop, Missing Ack, Address, Data, or Address and Data
Maximum Data Rate: 10 Mbps
SPI Data Trigger: 1 to 16 bytes of user-specified data
Trigger On: SS Active, MOSI, MISO, or MOSI & MISO
Maximum Data Rate: 50 Mbps
CAN Data Trigger: 1 to 8 bytes of user-specified data, including qualifiers of equal to (=), not equal
to (<>), less than (<), greater than (>), less than or equal to (<=), greater than or equal to (>=)
Trigger On: Start of Frame, Type of Frame, Identifier, Data, Identifier and Data, End of
Frame, Missing Ack, or Bit Stuffing Errors
Frame Type: Data, R emote, Error, Overload
Identifier: Standard (11 bit) and Extended (29 bit) identifiers
Maximum Data Rate: 1 Mbps
LIN
Identifier Trigger: 6 bits of user-specified data, equal to (=)
Data Trigger: 1 to 8 bytes of user-specified data, including qualifiers of equal to (=), not
equal to (<>), less than (<), greater than (>), less than or equal to (<=), greater than or
equal to (>=), inside range, or outside range
Error: Sync, Identifi er Parity, Checksum
Trigger On: S ync, Identifier, Data, Identifier & D ata, Wakeup Frame, Sleep Frame, or Error
Maximum Data Rate: 100 kbps
16 MSO4000B and DPO4000B Series Specifications and Performance Verification