EasyManuals Logo
Home>Tektronix>Test Equipment>P6015A

Tektronix P6015A User Manual

Tektronix P6015A
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
In Detail
P6015A Instruction Manual
117
DC Attenuation
1. Connect the P6015A compensation box to the oscilloscope.
2. Connect a BNC male-to-GR adapter to the calibration generator
amplitude output. Connect the P6015A ground lead clip to the
outer fin of the GR connector.
WARNING. The calibration generator produces dangerous voltages
during these procedures. To avoid electrical shock, be careful not to
touch the GR center conductor or exposed portions of the probe tip
while the generator is on.
3. Insert the tip of the P6015A into the GR center conductor.
4. Set the calibration generator to standard amplitude output of
50 Volts. This setting on a PG 506A produces a 1 kHz square
wave.
5. Set the oscilloscope time base to display 500 ms/div, and set the
vertical deflection to 10 mV/div (10 V/div when using the
readout option).
6. Center the waveform on the screen.
7. Adjust DC ATTEN (R9) for five divisions. Use the trailing portion
of the square wave if the leading corner is not flat.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix P6015A and is the answer not in the manual?

Tektronix P6015A Specifications

General IconGeneral
BrandTektronix
ModelP6015A
CategoryTest Equipment
LanguageEnglish

Related product manuals