Page #259 background image
Appendix B: Algorithms
Appendices
AĆ22
For example, if

is set directly, then

would not change even if
samples were out of range. However, if

was chosen using the %
choice from the Set Levels in % Units selection of the Measure menu, then

could give a “CLIPPING” warning.
NOTE
When measurements are displayed using Snapshot, out of range
warnings are NOT available. However, if you question the validity of
any measurement in the snapshot display, you can select and
display the measurement individually and then check for a warning
message.