EasyManua.ls Logo

Tektronix TDS 410A - Page 276

Tektronix TDS 410A
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
2–258
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:SLEwrate:POLarity
TDS 684A & 7XXA Only
Sets or queries the polarity for the main pulse slew rate trigger. This command is
equivalent to selecting Polarity in the Trigger menu with Type set to SlewRate.
Trigger
TRIGger:MAIn:PULse:SLEwrate:POLarity { EITher | NEGAtive |
POSITIVe }
TRIGger:MAIn:PULse:SLEwrate:POLarity?
<Space>
EITher
?
POSITIVe
NEGAtive
TRIGger :
MAIn
POLarity
:
PULse
:
:
SLEwrate
NEGAtive indicates that a pulse edge must traverse from the upper (most
positive) to lower (most negative) level for slew rate triggering to occur.
POSITIVe indicates that a pulse edge must traverse from the lower (most
negative) to higher (most positive) level for slew rate triggering to occur.
EITher indicates either NEGAtive or POSITIVe polarity.
TRIGGER:MAIN:PULSE:SLEwrate:POLARITY EITHER
specifies that the polarity of the slew rate can be either positive or negative.
TRIGger:MAIn:PULse:SLEwrate:SLEwrate? (Query Only)
TDS 684A & 7XXA Only
Returns the slew rate value. This is the
((Upper Threshold – Lower Threshold) / Delta Time)
The value is limited to the three most significant digits.
Trigger
Group
Syntax
Arguments
Examples
Group

Table of Contents

Related product manuals