EasyManua.ls Logo

Tektronix TDS 420A - Page 137

Tektronix TDS 420A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustment Procedures
TDS 420A, TDS 430A & TDS 460A Service Manual
5-11
H Press VERTICAL MENU.
H Push the main-menu button Bandwidth. Then push the side-menu
button 20 MHz.
H Press SHIFT.ThenpressACQUIRE MENU.
H Press the main-menu button Mode. Then press the side-menu
button Hi Res.
2. Compensate the Probe:
a. Locate the probe compensation hole in the side of the probe body (see
Figure 5--2).
b. Using the probe compensation tool, adjust the probe for best square
wave compensation (flat as possible across the top of the waveform).
(See Figure 5--3.)
Figure 5- 2: Performing Pr obe Compensation

Table of Contents

Other manuals for Tektronix TDS 420A

Related product manuals