EasyManuals Logo

Tektronix TDS 500C Performance Verification Manual

Tektronix TDS 500C
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
Typical Characteristics
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
2–29
Table 2–17: Typical characteristics — Triggering system (cont.)
Name Description
d
,
n
u
Pu
nd
r
, f
r
For vertical settings > 10 mV/div. and 3 1 V/div at the BNC input
Pu
Tr
r
n
The minimum pulse widths and rearm
Pulse class
Minimum pulse
width
Minimum re-arm width
widths and transition times
required for
Glitch 1 ns 2 ns + 5% of Glitch Width Setting
Pu
-Typ
r
r
n
.
Runt
2 ns 2 ns
Time-Qualified Runt 2 ns TDS 600B: 7 ns + 5% of Width Setting
TDS 700C: 8.5 ns + 5% of Width Setting
Width 1 ns 2 ns + 5% of Width Upper Limit Setting
Timeout 1 ns 2 ns + 5% of Width Upper Limit Setting
Slew Rate 600 ps
8
TDS 600B: 7 ns + 5% of Delta Time Setting
TDS 700C: 8.5 ns + 5% of Delta Time
Setting
Setup/Hold Time Violation Trigger,
Minimum Clock Pulse Widths
For vertical settings > 10 mV/div and 1 V/div at the BNC input,
the minimum requirements are:
Minimum Pulsewidth,
Clock High
Minimum Pulsewidth,
Clock Low
Users Hold Time + 2.5 ns
9
2 ns
Input Signal Sync Amplitude for Stable
Triggering, HDTV and FLEXFMT modes
(Option 05 Video Trigger)
All field selections: 0.6 division to 4 divisions
Jitter for HDTV mode
(Option 05 Video Trigger)
17 ns
p-p
Sync Width Flex Format and HDTV
modes (Option 05 Video Trigger)
min. 400 ns
Sync Duty Cycle, Flex Format and HDTV
modes (Option 05 Video Trigger)
min. 50 to 1
Hum Rejection
(Option 05 Video Trigger)
NTSC and PAL: –20 dB without any trigger spec deterioration. Triggering will continue
down to 0 dB with some performance deterioration.
1
The trigger position errors are typically less than the values given here. These values are for triggering signals having a
slew rate at the trigger point of 0.5 division/ns.
2
The waveform interval (WI) is the time between the samples in the waveform record. Also, see the footnote for the
characteristics Sample Rate Range or Interpolated Waveform Rates in Table 2–4, on page 2–10.
3
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on the
selected slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not
“roll” across the screen on successive acquisitions. The TRIG’D LED stays constantly lighted when the SEC/DIV setting is
2 ms or faster but may flash when the SEC/DIV setting is 10 ms or slower.
4
The minimum signal levels required for stable logic or pulse triggering of an acquisition, or for stable counting of a
DC-coupled, events-delay signal. Also, see the footnote for Sensitivity, Edge-Type Trigger, DC Coupled in this table. (Stable
counting of events is counting that misses no events and produces no extra, phantom events).

Table of Contents

Other manuals for Tektronix TDS 500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 500C and is the answer not in the manual?

Tektronix TDS 500C Specifications

General IconGeneral
BrandTektronix
ModelTDS 500C
CategoryTest Equipment
LanguageEnglish

Related product manuals