EasyManua.ls Logo

Tektronix TDS3000B Series - Signal Path Compensation and Time;Date Setup

Tektronix TDS3000B Series
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
1-2
TDS3000B Series Service Manual
Table 1- 1: Specifications (cont.)
Inputs
Input coupling DC, AC, or GND
Channel input remains terminated when using GND coupling.
Input impedance,
DC coupled
1MΩ ±1% in parallel with 13 pF ±2 pF, TekProbe compatible
50 Ī© ±1%; VSWR ≤ 1.5:1 from DC to 500 MHz, typical
Maximum voltage at Overvoltage category Maximum voltage
g
input BNC (1 MĪ©)
CAT I Environment
(refer to page 1--11)
150 V
RMS
(400 V
pk
)
CAT II Environment
(refer to page 1--11)
100 V
RMS
(400 V
pk
)
For steady-state sinusoidal waveforms, derate at 20 dB/decade above
200kHzto13V
pk
at 3 MHz and above.
Maximum voltage at
input BNC (50 Ī©)
5V
RMS
with peaks ≤±30 V
Maximum floating
voltage
0 V from chassis (BNC) ground to earth ground, or
30 V
RMS
(42 V
pk
) only under these conditions: no signal voltages
>30 V
RMS
(>42 V
pk
), all common leads connected to the same voltage,
no grounded peripherals attached
Channel-to-channel
crosstalk, typical
Measured on one channel, with test signal applied to another channel,
and with the same scale and coupling settings on each channel
,
y
p
Frequency range Crosstalk
≤ 100 MHz ≄ 100:1
≤ 200 MHz ≄ 50:1
≤ 300 MHz ≄ 50:1
≤ 400 MHz ≄ 30:1
≤ 500 MHz ≄ 30:1
≤ 600 MHz ≄ 30:1
Differential delay,
typical
100 ps between any two channels with the same scale and coupling
settings

Other manuals for Tektronix TDS3000B Series