Operating Information
TDS3000C Series S ervice Manual
1-21
Probe Compensation
Perform this adjustment to match your probe to the input channel. This should be
done whenever you attach a passive probe for the first time to any input channel.
1. Connect the oscilloscope probe to channel 1.
Attach the probe tip and reference lead to the
PROBE COMP connectors, t hen push AUTO-
SET.
If using the probe hook-ti p, e nsure a proper
connection by firmly twisting the tip onto the
probe.
CH 1
PROBE COMP
UTOSET
button
2. Check the shape of the displayed waveform.
Overcompensated
Compensated correctly
Undercompensated
3. If necessary, adjust your probe.
Repeat as necessary.
L
P3010 P6139A