Brief Procedures
2-10
TDS5000B Series Specifications and Performance Verification
H Rotating the horizontal SC ALE knob clockwise expands the waveform
on-screen (more horizontal divisions per waveform period), counter-
clockwise rotation contracts it, and returning the horizontal scale to
200 s/div returns the period to about five divisions.
H The horizontal POS ITION knob positions the signal left and right
on-screen when rotated.
6. Verify horizontal delay:
a. Center a rising edge on screen:
H Set the horizontal POS ITION knob so the rising edge, where the
waveform is triggered, lines up with the center horizontal graticule.
H Change the horizontal SCALE to 20 s/div. The rising edge of the
waveform should remain near the center graticule and the falling
edge should be off screen.
b. Turn on and set horizontal delay:
H Pull down the Horiz/Acq menu to select Horizontal/Acquisition
Setup. . . .
H Click the Horizontal tab in the control window that displays.
H Click the Delay Mode button to turn delay on.
H Double click the Horiz Delay control in the control window to
display the pop-up keypad. Click the keypad buttons to set the
horizontal delay to 500 s, and then click the ENTER key.
c. Verify the waveform: Verify that a falling edge of the waveform is within
a few divisions of center screen.
d. Adjust the horizontal delay: Rotate the upper multipurpose knob to
change the horizontal delay setting. Verify that the falling edge shifts
horizontally. Rotate the front-panel horizontal POSITION knob. Verify
that this knob has the same effect (it also adjusts delay, but only when
delay mode is on).
e. Verify the delay toggle function:
H Rotate the front-panel horizontal POSITION knob to center the
falling edge horizontally on the screen.
H Change the horizontal SCALE to 40 n s/div (50 ns/div for
TDS5054BE). The falling edge of the waveform should remain near
the center graticule. If not, readjust the delay setting to center the
falling edge.

Table of Contents

Other manuals for Tektronix TDS5104B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5104B and is the answer not in the manual?

Tektronix TDS5104B Specifications

General IconGeneral
BrandTektronix
ModelTDS5104B
CategoryTest Equipment
LanguageEnglish

Related product manuals