Creating and Using Math Waveforms
3- 218
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
H Nearest Side Lobe. This is the difference in magnitude between the spectral
lobe peak in the spectrum and the next side lobe that occurs due to energy
leakage. Different windows have different leakage characteristics. The more
narrow the resolution bandwidth of the window, the more leakage in the
spectrum.
H Zero Phase Reference. This is the position in the time domain gate that is the
reference point for phase in the output spectrum. That is, if a sine wave input
has its peak at the zero phase reference position, then it reads out as zero
phase in the spectrum. If the phase is to be correct when doing impulse
response testing, the impulse in the time domain must be located at this
position in the gate interval.
H Coefficients. These are used to generate the windows which are constructed
from a cosines series. For the Gaussian window the value of a is given
instead of a set of coef ficients. You can find descriptions of cosine series
windows in Handbook of Digital Signal Processing Engineering Applica-
tions by Elliot. ISBN 0--12--237075--9.

Table of Contents

Other manuals for Tektronix TDS7104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS7104 and is the answer not in the manual?

Tektronix TDS7104 Specifications

General IconGeneral
BrandTektronix
ModelTDS7104
CategoryTest Equipment
LanguageEnglish

Related product manuals