Performance Tests
TDS7104 & TDS7054 Service Manual
4-45
2. Confirm the input channels are within limits for analog bandwidth: Do the
following substeps test CH 1 first, skipping substeps a and b since CH 1
is already set up for testing from step 1.
a. Select an unchecked channel:
H From the tool bar, touch MEAS and then Clear to remove the
previous measurement.
H Press the Vertical button of the channel just confirmed to remove the
channel from the display.
H Press the front-panel Vertical button that corresponds to the channel
you are to confirm.
H Move the leveling output of the sine wave generator to the channel
you selected.
b. Match the trigger source to the channel selected: Press the Trigger
SOURCE button until the source that corresponds to the channel you are
to confirm is on.
c. Set its input impedance: From the button bar, touch the VERT button
and select the tab for the channel you are to confirm. Touch the
Termination 50 button.
d. Set the vertical scale: Set the vertical SCALE that corresponds to the
channel you are to confirm to one of the settings listed in Table 4--7 not
yet checked. (Start with the 100 mV setting).
e. Set the triggering coupling: Touch the Coupling DC button.

Table of Contents

Other manuals for Tektronix TDS7104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS7104 and is the answer not in the manual?

Tektronix TDS7104 Specifications

General IconGeneral
BrandTektronix
ModelTDS7104
CategoryTest Equipment
LanguageEnglish

Related product manuals