EasyManuals Logo

Tektronix WFM6120 Technical Reference

Tektronix WFM6120
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
Video and General Performance Verication Procedures
Composite Analog Field
Rate Tilt (Option CPS)
Apply a eld rat
e square wave and use cursors to measure the tilt.
1. Set the waveform monitor to the factory presets. (See page 60, Incoming
Inspect
ion Tests.)
2. Connect the precision video generator to one of the CMPST A input
c
onnectors and terminate the other connector with the precision terminator.
3. Set the precision generator for Field Square Wave.
4. Press the Input C button to select the Composite A input.
5. Press the FULL button to make the WFM tile the full screen display.
6. Press the CURSORS buttontoturnonthecursorcontrols.
7. UsethearrowkeystoselectVolt1 and Volt2 and adjust the cursors to be near
the top of the waveform using the GENERAL knob.
8. Use the vertical knob to move the top of the waveform to the zero graticule.
9. Press and hold the GAIN button, select Gain Settings, and then select x5.
10. Press the GAIN button to close the menu.
11. Press the LINE SEL button use the left and right arrow keys t o select ALL.
12. Use the GENERAL knob to set the line select to a line number between
72 and 202.
13. Use the arrow keys to select Vol t1 and Vo lt 2 and use cursors to measure the
tilt in the waveform using the GENERAL knob to adjust them.
14. Record the value in the test record.
Composite Analog Line
Rate tilt (Option CPS)
Apply a eld rate square wave and use cursors to measure the tilt.
1. Set the waveform monitor to the factory presets. (See page 60, Restore the
Factory Presets.)
2. Connect the precision video generator to one of the CMPST A input
connectors and terminate the other connector with the p recision terminator.
3. Set the precision generator for 100% Flat Field.
4. Press the Input C button to select the Composite A input.
5. Press the FULL button to make the WFM tile the full screen display.
6. Press the CURSORS buttontoturnonthecursorcontrols.
7. Use the arrow keys to select Volt1 or Volt2 and adjust cursors to be near the
top of the waveform using the GENERAL knob.
8. Use the VERTICAL position knob to move the top of the waveform to the
zero graticule.
Waveform Monitors Specications and P erformance Verication 113

Table of Contents

Other manuals for Tektronix WFM6120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WFM6120 and is the answer not in the manual?

Tektronix WFM6120 Specifications

General IconGeneral
BrandTektronix
ModelWFM6120
CategoryMonitor
LanguageEnglish

Related product manuals