Pathfinder DVL Guide April 2018
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 117
CZ Power Down ADCP
Purpose Tells the Pathfinder to power down.
Format CZ
Recommended Setting. Use as needed.
Description Sending the CZ command powers down the DVL and draws a minimal amount of current.
Pathfinder processing is interrupted and the DVL goes in the STANDBY mode (RAM is
maintained).
Example See below
>cz
Powering Down
A Hard Break will re-wake the unit. Note that this only works over serial, not Ethernet.
Only a Hard Break will wake up the DVL from sleep. Using a Software Break will not work.
Expert Control System Commands
This section lists the expert Control System commands. Commands that start with the # sign are consid-
ered “expert” commands.
>#c?
Available Commands:
#CT 0 ---------------------- Turnkey [0 = OFF, 1 = ON]
#C? ------------------------ Display #C-Command Menu