DATA OUTPUT
Axxx<cr> Output each marked station value
every xxx cycles, where xxx is any value
between 1 and 255.
CA<cr> Cancel the automatic output.
Mx<cr> Mark station x, where x is any number between 1 and 9, or A for Station
10.
Rx<cr> One shot reading of station x.
Ux<cr> Unmark station #x from automatic output.
DISPLAY
DLx<cr> Station x is displayed on the left side.
DRx<cr> Station x is displayed on the right side.
FA(cr) In conjunction with the "FI" command, will inhibit all front panel function,
except station select.
FE<cr> Enables all front panel functions, unless
inhibited by the rear panel switch.
FI<cr> Inhibits certain front panel functions.
UM<cr> Causes Stations to be displayed in millibars.
UP<cr> Causes stations to be displayed in Pascals.
UT<cr> Causes stations to display in traditional units.
GAS
GFx<cr> Clears Station x from gas operation.
GNx<cr> Sets Station x for gas (Argon) operation.
HOT CATHODE
AH<cr> Auto Mode For Hot Cathode Gauges.
EH<cr> Enable Hot Cathode Gauges (shutdown mode).
FF<cr> 3D Filament OFF.
FN<cr> 3D Filament ON.
GF<cr> 3D Degas OFF.
GN<cr> 3D Degas ON.
GNxxx<cr> 3D Degas ON for "xxx" minutes, 255 max.
SWx<cr> Make sensor switch setting equal "x".
LEAK RATE
ED<cr> End Leak Rate data stream (but not the test).
EL<cr> Stops Leak Rate test.
LR<cr> Starts Leak Rate test. Time = 0.
RL<cr> Read Leak Rate value periodically. "LR=nnnn"
where nnnn is the leak rate in microns/hr.
OUTPUT FORMATS
R1EL<cr> 0-10V of 1Es to be logarithmic.
R1En<cr> 0-10V of 1Es to be linear.