7.1 Controller Features
Temperature conditions are controlled with the standard profiling type Watlow 942 Controller. The alternate
Watlow F4 Controller may be used for more demanding applications. Both controllers are described below.
Watlow 942 Controller:
The Watlow 942 is a profiling type ¼ DIN controller with ramping and auto-tuning control. The following main
features are employed.
Single Channel
Profiling: Multiple Profiles, 24 Steps Max.
Automatic & Manual Control
1 Input: Universal Type (RTD is Std.)
2 Control Outputs: Time Proportioned, or On / Off Control
2 Auxiliary Outputs - Optional: 1 or 2 Events, or 1 or 2 Alarms, or 1Retransmit w/Alarm, or w/Event
Guaranteed Soak - Standard
Communications: Serial - Optional
Watlow F4S Controller:
The Watlow F4S is a profiling type ¼ DIN controller with ramping and auto-tuning control. The following main
features are employed.
Single Channel
Profiling: Forty Profiles, 256 Steps
Automatic & Manual Control
1 Input: Universal Type (RTD is Std.)
4 Digital Inputs
2 Auxiliary Analog Inputs - Optional
2 Control Outputs: Time Proportioned, On/Off, or 4-20 ma
2 Alarm Outputs
8 Digital Event Outputs
2 Retransmit Outputs - Optional
Communications: EIA-232 and EIA-485 with Modbus RTU protocol