INSTRUCTIONS DE REPARATION
REPARATURANLEITUNG
REPAIR INSTRUCTIONS
DOCUMENT No: 334.80051 V03
DATE :15/08/2002
After having switched on the M-H, the PC will communicate with the M-H and the following screen will appear :
Remark : If following screen does not appear after connecting the cable and switching on the M-H, please ensure that the
configuration of communication port is correct and that there is no cable problems.
2. Control / Correction of the slope
Press F1.
After having passed over the column reference mark and 'IG13 following screen will appear :
Note : If the IG13 are in frontal position you will correct x slope. If it is on lateral position you will introduce correction on ±Y.