EasyManua.ls Logo

TESTO 270 - Page 19

TESTO 270
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 First steps
19
If an incorrect PIN is entered, the instrument switches into
measuring mode.
- When the configuration menu is not locked, you can gain
access by setting the limit values.
2. Set the alarm limit values.
- Alarm lights up in the display.
> Set lower limit value (Alarm ): [] or [] and confirm with
[Hold].
- Alarm lights up in the display.
> Set upper limit value (Alarm ): [] or [] and confirm with
[Hold].
3. Carry out / do not carry out calibration.
- CAL and no or yes light up in the display.
> Activate / deactivate calibration / adjustment function (no is
the standard setting): [] or [] and confirm with [Hold].
- Selection of yes: The instrument can be calibrated /
adjusted, see Calibrating/adjusting the instrument, page 27
- Selection of no: No calibration / adjustment possible
4. Automatically hold readings.
- Auto-Hold and on or off light up in the display.
> Switch Auto-Hold on or off: [] or [] and confirm with
[Hold].
5. Automatically switch the instrument off.
- Auto-off and on or off light up in the display.
> Switch Auto-off on or off: [] or [] and confirm with
[Hold].
6. Set alarm indicator
- Alarm and on or off
light up in the display.
> Switch alarm indicator on or off: [] or [] and confirm with
[Hold].
7. Set the temperature unit.
- °C or °F lights up in the display.
> Set the temperature unit (°C/°F): [] or [] and confirm
with [Hold].
8. Enable / disable PIN.
- PIN and yes or no light up in the display.
> Activate PIN (yes) or deactivate it (no is the standard
setting).
9. Carry out reset.
- rst and yes or no light up.

Other manuals for TESTO 270

Related product manuals