7 Maintaining the product
27
With cold oil residues on the sensor
1. Immerse sensor in hot oil.
2. Allow sensor and probe shaft to cool until there is no longer a
risk of burns.
3. Clean sensor before the oil residues cool down.
Pos: 70 /TD/Produk t instand halten/tes to 275/Gehäuse reini gen (testo 275) @ 16\mod_1408 017834842_79.doc x @ 196931 @ 2 @ 1
7.3. Cleaning the housing
Requirement: The instrument is switched off.
CAUTION
Possible damage to housing!
> Do not use sharp-edged objects.
> Do not use any aggressive cleaning agents or solvents.
> Use mild household cleaning agents, standard household
washing-up liquids, water or soap suds.
> Clean housing with a damp cloth.
> Dry housing.
Pos: 71 /TD/Produk t instand halten/tes to 275/Koffer rei nigen (testo 275) @ 16\mod_14134 38648965_79.doc x @ 202997 @ 2 @ 1
7.4. Cleaning the plastic case
CAUTION
Possible damage on and in the plastic case!
> Do not use sharp-edged objects.
> Do not use any aggressive cleaning agents or solvents.
> Use mild household cleaning agents, standard household
washing-up liquids, water or soap suds.
> Clean plastic case with a damp cloth.
> Wipe plastic case with a dry cloth.
Pos: 72 /TD/Produk t instand halten/tes to 275/Instandhaltun g Gerät kalibrier en/justieren (testo275) @ 15\mod_1407495334611 _79.docx @ 195020 @ 255 @ 1
7.5. Calibrating/adjusting the instrument
You can check the accuracy of your instrument by carrying out a
comparative measurement with the Testo reference oil (calibrating).
If there is too great a difference between the reading and the
reference value, future instrument readings can be adjusted to the
reference value (adjusting).