EasyManua.ls Logo

TESTO 270 - Page 29

TESTO 270
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Maintaining the product
29
2. Go through the configuration mode with [Hold], until CAL and
yes or no light up in the display.
3. Switch on the calibration/adjustment function with [] or []
(yes).
4. Confirm with [Hold].
- OIL and CAL light up in the display.
There is a negative effect on the accuracy of the
calibration/adjustment when the reference oil bottle is held
by hand.
5. Immerse sensor in reference oil. Adhere to the immersion
depth!
- When alarm indicator is activated: the display lights up yellow.
6. Start calibration/adjustment process with [Hold]. For faster
recording of readings: move sensor in oil.
- The display lights up red.
- Measured TPM value and temperature are displayed.
- With a stable reading: Control keys are enabled, display lights
up green.
7. Compare the value shown in the display with the target value
indicated on the label of the reference oil bottle.
> If there is a deviation of > 1%, adjustment should be carried
out. Continue with step 8.
> If there is a deviation of </= 1%, no adjustment is required.
Continue with step 9.
Adjustment with reference oil impairs the accuracy by
0.5% TPM compared to the factory calibration.

Other manuals for TESTO 270

Related product manuals