7 Maintaining the product
48
7.7. Condensate container
The fill level of the condensate container can be read from the
markings on the condensate trap.
Draining the condensate container
The condensate consists of a weak mix of acids. Avoid skin
contact. Make sure that the condensate does not run over
the housing.
ATTENTION
Damage to the sensors and the flue gas pump due to
condensate entering the gas path!
> Do not empty the condensate container while the flue gas
pump is in operation.
1. Open the condensate outlet on the condensate trap: Pull out
the plug as far as it will go.
2. Let the condensate run out into a sink.
3. Wipe off any drops still on the condensate outlet with a cloth
and close the condensate outlet.
The condensate outlet must be completely closed
(marking), otherwise measuring errors could be caused by
infiltrated air.