9 Maintenance
41
9 Maintenance
9.1 Calibration
The testo 560i scale and the testo Smart Valve are supplied with a
factory calibration certificate as standard.
Recalibration once every 12 months is recommended in many
applications.
This can be carried out by Testo Industrial Services (TIS) or other
certified service providers.
Please contact testo for further information.
Do not use any aggressive cleaning agents or solvents! Mild household
cleaning agents and soap suds may be used.
If the instrument housing is dirty, clean it with a damp cloth.
Connect high pressure air and use the function to click 3 times the
button and the valve will open for 1 s. This can help to remove small
dust inside the valve.
9.3 Keeping connections clean
Keep screw connections clean and free of grease and other deposits;
clean with a damp cloth as required.
9.4 Removing oil residues
Carefully blow out oil residues in the valve block using compressed
air.
9.5 Ensuring measuring accuracy
Testo Customer Service will be happy to help you as required.
Check the instrument regularly for leaks. Keep to the permissible
pressure range!
Calibrate the instrument regularly (recommendation: once a year).
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com