116. Setting instrument
2
Setting ddegree oof eemission ((
ε
):
The configuration mode is open.
i Set value with / and confirm with .
º Continue with SELECTION OF ALARM VALUE TO BE MONITORED.
3
Selection oof aalarm vvalue tto bbe mmonitored ((
dtd, Ir, rH Surface
):
The configuration mode is open.
Select the desired option with / and confirm with .
º Continue with handling objective SETTING UPPER LIMIT VALUE (IR).
or
º Continue with handling objective SETTING LIMIT VALUE ΔTD
or
º Continue with handling objective SETTING LIMIT VALUE RHSI.
4
Setting uupper llimit vvalue ((
Ir
):
The configuration mode is open.
i Select upper limit value with / and confirm with .
If the upper limit value is exceeded, the max. value is displayed in the top
line (symbol for exceeded upper limit value appears). The IR temperature
is displayed as the main value. The corresponding limit/alarm value is
displayed in the lower line. The alarm symbol appears. If an acoustic alarm
has been selected, it sounds. If the value drops back under the limit value,
the instrument returns to the display previously shown.
º Continue with SETTING LOWER LIMIT VALUE (IR).
5
Setting llower llimit vvalue ((
Ir
):
The configuration mode is open.
i Select lower limit value with / and confirm with .
If the lower limit value is exceeded, the max. value is displayed in the top
line (symbol for exceeded lower limit value appears). The IR temperature
is displayed as the main value. The corresponding limit/alarm value is
displayed in the lower line. The alarm symbol appears. If an acoustic alarm
has been selected, it sounds. If the value climbs back over the limit value,
the instrument returns to the display previously shown.
º Continue with handling objective AUDIBLE ALARM ON/OFF.