7 Maintaining the product
165
Saveris T4E Ethernet probe
4-channel Ethernet probe with 4 external TC probe connections
Feature Values
Probe type TC
Measuring range
TC type S
TC type T
TC type J
TC type K
0 to +1760 °C
-200 to +400 °C
-100 to +750 °C
-195 to +1350 °C
Accuracy ±0.5 °C or 0.5% of measured value
Resolution 0.1 °C/TC type S 1 °C
Connection
Mini-DIN service interface is accessible
externally
4 TCs via TC socket,
max. difference in potential 50 V
Weight Approx. 220 g
The technical data refer to the probe in a stable,
stationary operating mode. To receive a stable
measurement the probe must be in operation for one to
two hours.
The usage of isolate thermocouples is recommended
for supplying the Ethernet probe. Otherwise
discrepancies in readings of up to 0.6 °C may occur due
to leakage currents.