Power
line
conditioning
accessories
Electrical
service
specifications
16 Nicolet
Uninterruptable power supplies (UPS) are available from Thermo
Nicolet. A UPS reduces the odds
of
a system shutdown
if
power is
lost elsewhere in the building. Power line conditioners (which ensure
that
your
service
is
free
from
sags, surges or other line disturbances)
also are available in the U.S.A. from Thermo Nicolet for 120-volt
operation. Line conditioners for 220-volt operation can be purchased
locally. Contact Thermo Nicolet Customer Support for information
about power conditioners and UPS.
The following table lists the specifications for electrical service.
Contact your local Thermo Nicolet service representative
if
you have
questions about the requirements.
If
you are not sure that your power
lines meet these requirements, contact Thermo Nicolet Customer
Support for information about power audits.
Requirement Specification
AC
input 100 to 240 VAC
line frequency 50-60 Hz
current
15
A (120-volt operation)
7.5 A (220-volt operation)
line disturbances Sags, surges or other line disturbances must not
exceed 10%
of
input voltage (even for a half
cycle).
noise less than 2 volts ( common mode)
less than 20 volts (normal mode)
\_
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the THERMO NICOLET Avatar Series and is the answer not in the manual?

THERMO NICOLET Avatar Series Specifications

General IconGeneral
BrandTHERMO NICOLET
ModelAvatar Series
CategoryMeasuring Instruments
LanguageEnglish