Purge
gas
generators
Whatman Model Numbers:
Thermo Nicolet part number
nominal input line voltage
dew point
maximum dry (outlet) air flow
rate for specified dew points*
inlet pressure 8.50 atm (125
psig)
inlet pressure 4.08 atm (60 psig)
air consumption for
regeneration**
minimum inlet air pressure
CO2 concentration
maximum inlet air temp.***
inlet/outlet port size
shipping weight
If your facility does not have a source
of
clean, dry compressed air
or nitrogen for system purge, we recommend that you consider using
a purge gas generator. The generator cleans and dries the air supplied
by your air compressor and provides it to the spectrometer. (For
facilities without an air compressor, a complete dry-air generating
system is available; see the next section.)
The spectrometer works well with the Whatman® (Balston®) purge
gas generator models shown at the top
of
the following table, which
lists part numbers and specifications. The generator supply voltage and
frequency are not adjustable, so be sure to order the correct items for
your local utility service. The power consumption
of
the generators is
listed earlier in the section called "Power consumption."
75-45 75-45EU
75-52
75-
52EU
869-050300 869-050400 869-050500
869-050600
120 volts AC 240 volts AC 120 volts AC 240 volts AC
-73°C (-100°F)
-73°C (-100°F)
-73°C (-100°F) -73°C (-100°F)
17
liters/min
17
liters/min 34 liters/min 34 liters/min
(36 scth)
(36 scth) (72 scth) (72 scth)
9 liters/min 9 liters/min 17 liters/min
17
liters/min
(18 scth)
(18 scth) (36 scth) (36 scth)
14 liters/min
14 liters/min 28 liters/min 28 liters/min
(30 scth)
(30 scth) (60 scth)
(60 scth)
4.
08
atm (60 psig)
4.08 atm (60 psig) 4.08 atm (60 psig)
4.08 atm (60
psig)
< 1 ppm
< 1 ppm < 1 ppm < 1 ppm
25
°C (78°F)
25
°C (78°F)
25
°C (78°F) 25°C (78°F)
1/4 inch NPT
1/4 inch NPT 1/4 inch NPT 1/4 inch NPT
(female)
(female) (female)
(female)
11
kg (25 lb)
11
kg (25 lb)
18kg(40lb
)
18
kg (40 lb)
*
Dew point will be lower than specified at lower
ai
r
fl
ow.
**
Total air required = air loss + process demand (up to maximum dry air flow rate).
***
Ou
tl
et
dew point will increase at
hi
gher inlet compressed
ai
r temperatures.
22
Nicolet
'---
'---
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the THERMO NICOLET Avatar Series and is the answer not in the manual?

THERMO NICOLET Avatar Series Specifications

General IconGeneral
BrandTHERMO NICOLET
ModelAvatar Series
CategoryMeasuring Instruments
LanguageEnglish