Follow the recommendations below to prepare your site for installing
the Thermo Nicolet Liquid Analysis System.
We
highly
recommend
dry-air
purge.
See
"Purge gas"
for
more
information.
The configured system requires about 2 meters (6 feet)
of
spectrometer space. The autosampler must be located on the right
side
of
the spectrometer.
A wash solvent reservoir containing the appropriate solvent must
be located behind the autosampler. The reservoir is not included
with the system.
The waste reservoir (provided) must be located below the system
for gravity-fed waste containment.
Be sure to have a sample material that is appropriate for
performing a pathlength calibration
of
the sample cell. For
example, the material required for used lubricating oil analysis is
99% HPLC grade or purer heptane. You will need 2 to 8 liters
of
this material per year, depending on the frequency
of
calibration.
Be prepared to supply some representative samples for verifying
that the system is working properly at the time
of
installation.
'---
'---
'--
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the THERMO NICOLET Avatar Series and is the answer not in the manual?

THERMO NICOLET Avatar Series Specifications

General IconGeneral
BrandTHERMO NICOLET
ModelAvatar Series
CategoryMeasuring Instruments
LanguageEnglish