Follow the recommendations below to prepare your site for installing
the Thermo Nicolet Liquid Analysis System.
•
We
highly
recommend
dry-air
purge.
See
"Purge gas"
for
more
information.
• The configured system requires about 2 meters (6 feet)
of
spectrometer space. The autosampler must be located on the right
side
of
the spectrometer.
• A wash solvent reservoir containing the appropriate solvent must
be located behind the autosampler. The reservoir is not included
with the system.
• The waste reservoir (provided) must be located below the system
for gravity-fed waste containment.
• Be sure to have a sample material that is appropriate for
performing a pathlength calibration
of
the sample cell. For
example, the material required for used lubricating oil analysis is
99% HPLC grade or purer heptane. You will need 2 to 8 liters
of
this material per year, depending on the frequency
of
calibration.
• Be prepared to supply some representative samples for verifying
that the system is working properly at the time
of
installation.
'---
'---
'--