EasyManua.ls Logo

Thermo Scientific Prima PRO - Page 52

Thermo Scientific Prima PRO
203 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Startup & Shutdown
Procedure
Thermo Fisher Scientific Prima PRO & Sentinel PRO Mass Spectrometers User Guide 7-3
which shows the alarm status of each of the instrument
parameters.
Figure 7–2. Instrument Status window
A click on a particular parameter status light will show the
current value, alarm limits, and alarm acknowledgement button
(Figure 7–3). If all parameters are within limits but the status
light is red, clicking on the Acknowledge All button will change
the instrument status light to green.
Wait until all parameters are within limits (the status light is
green), before starting scheduled analysis.
Figure 7–3. Parameter status window

Table of Contents