EasyManua.ls Logo

THORLABS PM100A - Page 52

THORLABS PM100A
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.5 SCPI Commands
50
6.5.2.4 INPut subsystem commands
Command
Description
INPut
[:PDIode]
:FILTer
[:LPASs]
[STATe] {OFF|0|ON|1}
[STATe]?
Sets the bandwidth of the photodiode input stage
Queries the bandwidth of the photodiode inut stage
:THERmopile
:ACCelerator
[STATe] {OFF|0|ON|1}
[STATe]?
:TAU {MINimum|
MAXimum|DEFault|
<numeric_value>[s]}
:TAU? [{MINimum|
MAXimum|DEFault}]
Sets the thermopile accelerator state
Queries the thermopile accelerator state
Sets thermopile time constant
(0-63%)
in s
Queries the thermopile time constant
(0-63%)
in s
:ADAPter
[:TYPE] {PHOTodiode|
THERmal|PYRo}
[:TYPE]?
Sets default sensor adapter type
Queries default sensor adapter type
Commands for PM100D / PM100USB only

Table of Contents

Related product manuals