48
man_vd6s2-de-130515
6 Maintenance and Service
Danger of possibly contaminated parts!
Contaminated parts can cause personal injuries. Inform your-
self regarding possible contamination before you start working.
Be sure to follow the relevant instructions and take care of
necessary protective measures.
The unit requires no maintenance. External dirt and soiling can be removed by a
damp cloth.
Should a defect or damage occur on the VD6, please return the instrument for
repair, enclosing a contamination declaration.
The unit is not planned for customer repair!
Error messages and malfunction
"ur" measurement under range
"or" measurement over range
"Err1" no transducer connected
wrong transducer type
defective transducer
check connectors and
cables
select correct type (see
chapter 5.1)
send transducer for repair
"Err4" memory error
send VD6 for repair