5.3 Recorder Output
The analog signal of the VD6 recorder output (see chapter 3.6) can be config-
ured as 0-1V or 0-10V output.
Setting 0 – 1.0 V:
VD6 with transducer type 1 / VSC43MA4:
0 – 1 V, linear;
0.282 V corresponds to 1 mbar; 1.408 V corresponds to 1400 mbar
VD6 with transducer type 3 / VSP53MA4:
0 – 1.0 V, nonlinear according to characteristic curve (see below)
0.200 V corresponds to 0.001 mbar; 1.000 V corresponds to 1000 mbar
Pressure
(mbar)
Output (mV)
Pressure
(mbar)
Output
(mV)
0.001 200 0.80 610.1
0.002 203.4 1.0 647.2
0.004 208.4 2.0 755.8
0.006 213.5 4.0 849.3
0.008 217.7 6.0 892.2
0.010 221.9 8.0 914.9
0.020 239.6 10 930.9
0.040 270.7 20 963.8
0.060 296.0 40 981.5
0.080 317.1 60 986.5
0.10 335.6 80 990.7
0.20 403.8 100 992.4
0.40 498.9 1000 1000
0.60 562.9
VD6 with transducer type 7 / VCC200MA4:
0 – 1 V, linear;
0.5 mV corresponds to 0,1 mbar; 1.000 V corresponds to 200 mbar
Setting 0 – 10 V:
VD6 with transducer type 1/ VSC43MA4:
0 – 10 V, linear;
5 mV / mbar
VD6 with transducer type 3 / VSP53MA4:
0 – 10 V, logarithmic; (1V per Decade)
2.5 V corresponds to 1x10
-3
mbar;
8.5 V corresponds to 1000 mbar
V
out
(V) = log (p(mbar)) + 5.5
p (mbar) = 10
(Vout(V) - 5.5)
V
out
(V) =
0.
005 p(mbar)
p (mbar) = V
out
(V) / 0.005