EasyManuals Logo

Titan Logix Rack Control Module Application Note

Titan Logix Rack Control Module
11 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
Application Note
TPM 060, Rev 1.0
September 2018
Page
8
of
11
5. As an additional test, if possible, add fluid or drain the fluid from the tank and observe the display on
the Finch II. The display should change correspondingly – decreasing when draining and increasing
when adding fluid to the tank.
6. Retrofit Procedures with empty tank
6.1 Retrieve the TD80 strap table
1. Ensure power is applied to the Finch II/TD80 system.
2. Take note if there is a sump and/or riser on the tank. Record their corresponding values.
3. Read the strap table off the TD80 using SensorLink utility program.
a. Information relating to the use of the SensorLink utility program can be found in TPM 058
SensorLink User Manual.
4. Save the TD80 strap table as a file for future reference.
5. Remove power from the Finch II/TD80 system.
6.2 Exchanging the transmitter
1. An external Dual-Rod probe is required for this procedure. Place and secure the shorting block of the
Dual-Rod probe at the location about 32 inches or 81cm from the fiducial.

Other manuals for Titan Logix Rack Control Module

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Titan Logix Rack Control Module and is the answer not in the manual?

Titan Logix Rack Control Module Specifications

General IconGeneral
BrandTitan Logix
ModelRack Control Module
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals