16/35
7.5. Example of communication
When set to the 50.00 for Higher and 40.00 for lower limit torques
※ Adjust the decimal point depend on the model and unit of measure.
Example 1) Set to 20 Nm for higher and 15 Nm for lower limit for CEM20N3X10D-G-BTA
Send a command the “AT037,20.00,15.00CRLF”
Example 2) Set to 90 Nm for higher, 80 Nm for lower limit for CEM100N3X15D-G-BTA
Send a command the “AT037,090.0,080.0CRLF”
Example 3) Set to 600 Nm for higher, 500 Nm for lower limit for CEM850N3X32D-G-BTA
Send a command the “AT037,0600.,0500.CRLF”
When set to the 15 degrees for double tightening detection, 30 degrees for lower and
100 degrees for higher limit angles
External device
A T 0 3 7 , 5 0 . 0 0 , 4 0 . 0 0
CR LF
CEM3-BTA
R E 0 0 3 , O K
CR LF
●
Hi/Lo limit torques
●
Response command
●
Double tightening
detection angle and
Hi/ Lo limit angles
●
Response command
External device CEM3-BTA
R E 0 0 3 , O K
CR LF
A T 0 4 6 , 0 1 5 , 0 3 0 , 1 0 0
CR LF