1-8
Be sure to turn off the power of this instrument and all instrumens which
are connected to it, when connecting a cable from an external device to
the connector on the instrument. Failure may occur.
1.1.5 External devices or network
When operating this instrument connected to other devices, only use
devices that satisfy IEC60601-1 or equivalent safety requirements, or that
conform to IEC60950 and whose power source is insulated with an
isolation transformer, in order to satisfy the relevant safety requirements
for medical equipment.
In case of setting the network, be sure to do it after agreement of the
network administrator.
For LAN connection, be sure to connect the instrument via a hub. The
instrument will not operate correctly if connected directly.
In cases where the instrument is connected to a network, there may be a
risk to the patient, operator or third party. Changing network may
present new risks. Be sure to confirm safety when you connect to, or
change a network.
1.1.6 Transportation and packing
The instrument needs to be protected against vibration and impact when
transporting it in a car or vehicle. Be sure to use the packaging box of
this instrument.
Keep the box and packing materials for use when moving or transporting
the instrument.